Mobile phase enters the 10-mm LightPipe flow cell through a port in the bottom of the flow
cell and exits through a port in the front of the flow cell as shown in
. Directing the
flow upward helps prevent the entrapment of air bubbles.
Figure 4.
Schematic of 10-mm LightPipe flow cell
Sensitive optical fibers are exposed at both ends of the flow cell (see
). Thermo Fisher
Scientific ships the LightPipe flow cell with end caps to protect these optical fibers (see
). Do not remove the protective end caps until you install the LightPipe flow cell.
Replace the protective end caps for storage if you remove the LightPipe flow cell from the
detector.
Figure 5.
End of LightPipe flow cell with a view of the sensitive optical fiber and the inlet port
Figure 6.
LightPipe flow cell with protective end caps
Inlet
Outlet
Alignment slot
CAUTION
Do
not
touch the ends of the LightPipe flow cell. Doing so could damage the
sensitive optical fibers. If you must grasp the ends of the LightPipe flow cell, wear clean,
talc-free gloves.
Optical fiber
Inlet port
10-mm LightPipe flow cell
50-mm LightPipe flow cell
Protective end caps for storage