Routine Maintenance
Replacing the Lamps
94
Surveyor PDA Plus Detector Hardware Manual
Thermo Scientific
9. To remove the deuterium lamp assembly:
a. Gently pull its connector free from the contact pins on the lamp power supply board.
As
shows, the connector for the deuterium lamp has three leads.
b. Loosen (but do not remove) the two Phillips head screws by approximately four
turns, lift and gently twist the mounting flange to free the lamp (see
).
Figure 81.
Lamp connections
Note
Do
not
try to remove the lamps from their mounting assemblies. Remove and
replace a lamp, its mount, cable, and connector as an assembly as described in the
following steps.
IMPORTANT
The surfaces of both lamps must be free of fingerprints and smudges,
which cause performance problems. For this reason, wear clean, talc-free gloves when
you handle the lamps. If either lamp requires cleaning, clean with a lint-free lens
paper moistened with methanol or isopropanol before replacing the lamp cover.
Deuterium lamp
connector
Tungsten lamp
connector