4
Start-up
Gas connection
4-10
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Gas connection
Installing gas pressure hoses
Figure 4-10.
Installing gas pressure hoses
The gas supply from the gas supply system to the device is established using the supplied flexible gas
pressure hoses:
1. Connect the gas pressure hose to the sleeve of the gas supply system.
2. Remove the protective cap (3/
) from the sterile filter.
3. Slide the hose clamp (1/
) onto the gas pressure hose (2/
) and connect hose
to the sleeve (4/
) of the sterile filter (5/
).
Note
Gas quality:
The gases must have one of the following qualities:
• Purity 99.5 % min
• medical gas quality.
CAUTION
Overpressure!
The operating pressure of the gas applied to the device must not exceed 1 bar (14.5 psi). If the gas is
supplied at a higher pressure, the valves integral to the device may not close correctly and the gas
supply control may be impaired.
Set the gas supply to a range between 0.8 bar (11.6 psi) min. and 1.0 bar (14.5 psi) max. and make
sure that this pressure setting cannot be changed!
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...