Handling and control
User configuration
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
6-35
• The selection dialog shown in
3. Select the menu command Gas tight screen.
is displayed.
Figure 6-39.
Gas tight screen setting
1. To toggle between two options:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the
Enter
key.
• The system returns to the
Options
menu.
Activating/de-activating the water level sensor
For incubation operation with ambient humidity or if the auto-start routine is to be run dry (without
water), the water level sensor can be switched off. This prevents alarm messages for the water level
sensor by the device-integral monitoring system from being issued:
Activating/de-activating the water level sensor
1. Press the
Menu
key.
2. Select the menu command
Options
.
• The selection dialog shown in
3. Select the menu command Water level sensor.
is displayed.
WARNING
When the water level sensor is de-activated, the steri-run routine can be started
despite of the presence of a water fill in the water reservoir. This is improper usage and may destroy
the fan motor.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...