8
Cleaning and disinfection
Steri-run decontamination routine
8-10
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
• Start steri-run.
• The run of the steri-run decontamination routine starts.
• While the steri-run decontamination routine is running, the display shows the current status and
outputs the following information:
• temperature,
• start time,
• phase,
• remaining run time.
Figure 8-6.
activate steri-run
cancel steri-run
The steri-run decontamination routine can be interrupted at any time.
1. Cancel steri-run:
• Press the
Stop
key.
• If the
Stop
key in the status display is depressed, the steri-run stop dialog box is displayed for a
safety scan. The routine can now be permanently cancelled or resumed.
2. To stop steri-run:
• Press the
End
key [3].
• The error message is displayed.
• If the error message is confirmed, the system returns to the main menu.
3. To resume steri-run:
• Press the
Back
key.
• The system returns to the status display, and the decontamination routine is resumed.
4. To interrupt Steri-Run from the status display:
• Press the
Stop
key.
• As a safety scan, the steri-run - Stop dialog window appears. Proceed with work step 2 (see section
above).
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...