8
Cleaning and disinfection
Wipe/spray disinfection
8-4
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Predisinfection:
1. Spray disinfectant onto the surfaces of the workspace and of the accessories and wipe the surfaces
clean.
2. Let the disinfectant work on the surfaces/internals as detailed in the manufacturer’s instructions.
Removing accessories and shelf system:
1. Remove the shelves, then remove the entire shelf system from the workspace. The installation and
removal of the shelving system are described in the section
.
2. Remove the airbox and the HEPA-filter. The installation and removal of the airbox and the
HEPA-filter are described in section
.
CAUTION
Health hazard!
The surfaces of the workspace may be contaminated. Contact with contaminated cleaning liquids
may cause infections. Disinfectants may contain harmful substances.
When cleaning and disinfecting, always observe the safety instructions and hygiene regulations!
• Wear safety gloves.
• Wear safety goggles.
• Wear mouth and respiratory system protection gear to protect your mucous membranes.
• Observe the safety instructions of the disinfectant’s manufacturer and the hygiene supervisor.
CAUTION
Moisture-sensitive components!
Do not spray the CO
2
sensor and the
2
-/N
2
sensor behind the air duct with disinfectant.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...