8
Cleaning and disinfection
Steri-run decontamination routine
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
8-11
Steri-run interruption due to error
If an error occurs while the decontamination routine is run, an error message is displayed and the
following actions are initiated:
• The decontamination routine automatically changes to the cool-down phase,
• the audible alarm sounds.
1. To confirm the audible alarm:
• Press any position on the display.
• The audible alarm is switched off. The
End
key is displayed. If the decontamination routine is not
cancelled after this, cool-down to the set temperature occurs.
2. Cancel steri-run:
• Press the
End
key [3].
• The error message is displayed.
• If the error message is confirmed, the system returns to the main menu.
Figure 8-7.
Interrupting / cancelling steri-run