Annex: Data communication
Data logger structure
12-12
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Repeated output of the most recently read data, this command
does not move the reading pointer. This command can be used to avoid data loss
after
a communication error has occurred
Each query command is responded with up to 7 data sets that follow one another without a separator.
These data sets consist of 16 bytes and are encrypted into 32 ASCII characters prior to transmission.
For example, the byte 0x23 is converted into the ASCII characters:
0x32 (“2“) and 0x33 (“3“).
Therefore, a response consists of up to 7 X 16 = 112 bytes, i.e. 224 ASCII characters.
The date and time (without seconds), the device status and the type of data logger entry are always
transmitted in a data set (byte 0-7, and ASCII characters 0-15).
Also, depending on the entry, the current actual values or set values of the control loops or other
parameters can be entered (byte 8-15 or ASCII character 16-31).
Example of a response:
!:2400:e0:010b01060f3700000177002800d40000110b01060f3800000172003200d20352... ...:80
First data set
!:2400:e0:010b01060f3700000177002800d4000011
(consisting of 32 byte ASCII characters)
Second data set
0b01060f3800000172003200d20352... ...:80
(beginning of the second data set after 32 bytes of the first data set)
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...