6
Handling and control
User configuration
6-30
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
The setting controls the logging cycle in sections of seconds during which the control loop measured
values are logged during the operation of the device and displayed by the trend display (
).
The settings can be made within the value range of 10 seconds to 3600 seconds.
1. To increase the value:
• Press the + key.
2. To reduce the value:
• Press the - key.
3. The value change appears in the display. The
New
message indicates that the selection has been
changed, but not saved yet.
4. To accept and save the change:
• Press the
Enter
key.
• The system returns to the
Event logging
menu.
Displaying the error table
The error table lists the errors detected by the device-integral monitoring system in descending
chronological order. The most recently detected error is listed at the top position of 22 possible entries.
An entry consists of the control loop in which the error occurred, the date, the time, and an error
description. The error table can be displayed but not edited. If the event display consists of two pages,
the user can browse through the list. The status indicator
001/002 or 002/002
indicates which page of
the two pages is currently being displayed.
Displaying the error table
1. Press the
Menu
key.
2. Select the menu command Event logging.
• The selection dialog shown in
3. Select the menu command Event display.
Logging cycle
Maximum period displayed
10 s
22.5 hours
30 s
2.8 days
60 s
5.6 days
120 s
11.2 days
180 s
16.8 days
3600 s
336 days
Note
Event logger interval time:
The logging cycle time does not affect the entries of the error table.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...