Factory presettings of the iCan™ Touchscreen controls
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
6-7
Factory presettings of the iCan™ Touchscreen controls
Upon delivery of the device, the following set values have been preset:
• Temperature: 37 °C
• CO
2
-content: 5,0 %
• O
2
-content (optional): 21,0 %
Heat-up phase of the control loop sensors
When the device has been switched on, the control loop sensors pass through a heat-up phase of
varying duration during the start process:
1. Temperature control loop: 10 s
2. CO
2
sensor heat-up period (TCD and IR): 5 min
3. O
2
control loop: 5 min
Figure 6-6.
Heat-up phase display
The start process is indicated by an audible signal. During the heat-up phase, the displays show dots
(...) instead of values:
• Temperature display,
• CO
2
display and
• O
2
display field
Note
CO
2
-/O
2
control:
Since the CO
2
concentration of the air is nearly 0%, the CO
2
control and the control loop error
monitoring system are disabled at a set value of 0%.
Since the O
2
concentration of the air is 21%, the O
2
control and the control loop error monitoring
system are disabled at a set value of 21%.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...