8
Cleaning and disinfection
Wipe/spray disinfection
8-6
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Figure 8-3.
Air duct
5. Unhinge the rear part of the air duct (2/
) from the rear wall and remove it.
6. Remove pre-filter (9/
).
The pre-filter is autoclavable.
7. Remove the cover panel from the water reservoir.
Cleaning the workspace and accessories:
1. Thoroughly remove dirt residues and deposits using a solution of tepid water and dishwashing
agent.
2. Wipe the surfaces clean using a clean cloth and plenty of clear water.
3. Remove the cleaning liquid from the water reservoir and wipe all surfaces of the workspace
thoroughly dry.
4. Wipe accessories thoroughly dry.
Final disinfection:
1. Spray the surfaces of the workspace, the shelving system and parts removed in previous steps with
disinfectant one more time and wipe dry.
2. Allow the disinfectant to react as specified by the manufacturer.
3. Reinstall the shelf system and the removed components.
F
E
D
C
A
1
2
B
G
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...