Annex: Data communication
Error memory structure
12-10
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Overview of the possible error messages in hex coding
Overview of the possible error messages in bit coding
General device status, temperature and CO
2
control loop:
Hex Code
Description / Type
0x00
Temperature control loop
0x01
CO
2 control loop
0x02
O
2 control loop
0x07
Water level
0x08
General device status
Bit
General device status
0x0002
Device door open too long
0x0004
Display not communicating
0x0008
Implausible mainboard parameter (EEPROM defective)
0x0010
Data logger defective (unit can continue to function)
0x0020
Error for disinfection / steri-run
0x0040
Power-down during steri-run
0x0080
Error auto-start
0x0100
ADC-test failed
0x0400
Fan error
0x1000
IR sensor system replaced (info)
0x2000
auto-start active (info)
0x4000
Disinfection active (info)
0x8000
Device error has occurred (info)
Bit
Error status of temperature control circuit
0x0001
Sensor breakage
0x0002
Actual value high
0x0004
Actual value low
0x0008
Actual value not plausible
0x0010
Calibration values too high/low
Bit
CO
2
control loop error status
0x0001
Sensor breakage
0x0002
Actual value high
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...