Annex: Data communication
Overview of incubator parameters (addresses 2xxx)
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
12-7
Reading internal function parameters
*1)
See table with notes on disinfection and autostart run status.
*2)
Two bytes each per level.
*3)
Cylinder A active (0x01), cylinder B active (0x02), pressure in cylinder A OK (0x10), pressure in cylinder B
OK (0x20).
*4)
For other Informations on the Error Log, see Chapter 13.5.
*5)
Place the reading pointer on the first entry, read up to 7 entries max.
*6)
Send the next 7 entries. Set reading pointer automatically to next new entry, read up to 7 entries
max.
*7)
Re-send entries of last telegram. Can be used after communication error has occurred.
Address
Description
Remarks
2100
Status of run
*1)
and
Remaining run time
[hours:minutes]
Disinfection, date
and time of last start
25 bytes / decimal value
with the format
xx;+xxx:xx;xx.xx.xx;xx:xx
2105
Status of run
*1)
, current
CO
2
waiting period
[minutes:seconds]
auto-start, Date and time of last
start
25 bytes / decimal value
with the format
xx;xx.x;+xxx:xx;xx.xx.xx;xx:xx
2140
Status of as cylinder changeover
switch
Read CO
2
*3)
2 bytes / hexadecimal
with the format xx
2141
Status of as cylinder changeover
switch
Read O
2
*3)
2 bytes / hexadecimal
with the format xx
2300
Read the Error Log
(current errors)
*4)
Up to 241 bytes / hexadecimal value
Format, see section on this topic
2301
Read the Error Log
(older errors)
*4)
Up to 241 bytes / hexadecimal value
Format, see section on this topic
2400
Query (Start) of data stored in the
data logger
*5)
Up to 224 bytes / hexadecimal value
Format, see section on this topic
2401
Query of other data stored in the
data logger
*6)
224 bytes / hexadecimal value
Format, see section on this topic
2402
(Repeat) query for last data logger
query
*7)
224 bytes / hexadecimal value
Format, see section on this topic
2410
Read out writing cycle of the data
logger in hours/minutes/seconds
8 bytes / decimal value with the
format xx:xx:xx
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...