3
Description of the device
Safety devices STERI-CYCLE i160 / i250
3-6
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
5. Gas pressure hose for looping-through the gas supply
6. Diagram: CO
2
and O
2
/N
2
gas connection with optional CO
2
and O
2
gas monitoring
7.
S
TERI-CYCLE
i160
/
S
TERI-CYCLE
i250
with looped-through gas supply; therefore the device must only
be equipped with combined gas connection (optional) CO
2
and O
2
/N
2
.
Safety devices
S
TERI-CYCLE
i160
/
i250
The device is equipped with the following safety devices:
• A door switch interrupts the CO
2
/O
2
/N
2
supply and the workspace heating when the glass door is
opened.
• When operated with the gas monitoring option, a gas monitor switches the gas supply to a full gas
tank.
• An overtemperature protection protects the samples from harmful overheating in case of failures.
• A pressure compensation opening ensures pressure compensation in the device workspace.
• The alarm relay circuit uses audible and visual alarms to indicate errors during operation.
Workspace atmosphere
In the workspace of the incubator, the particular physiological ambient conditions for the preparation
and cultivation of cell and tissue cultures are simulated. The workspace atmosphere is determined by
the following factors:
• Temperature
• Relative humidity
• CO
2
concentration
• O
2
concentration (optional)
Temperature:
To ensure undisturbed operation, the temperature in the operating room must be at least 18 °C and the
incubation temperature must exceed this temperature by at least 3 °C.
The heating system controls the incubation temperature from this temperature threshold up to 55 °C.
By heating the inner compartment with independent heater circuits and by separately heating the outer
door in addition, condensation is prevented on the sidewalls and the ceiling of the workspace and the
glass door or gas-tight screen.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...