8
Cleaning and disinfection
Decontamination procedures
8-2
S
TERI-CYCLE
i160
/
i250
Thermo Scientific
Decontamination procedures
The operator must prepare hygiene regulations for the decontamination of the device in accordance
with the application of the device.
The following disinfection procedures are compatible with the device:
Wipe/spray disinfection:
The wipe/spray disinfection is used as the standardized manual disinfection procedure for the device
and for all accessories.
Steri-run decontamination routine:
The steri-run decontamination routine uses an automated program cycle to decontaminate the
complete workspace including the shelf system and the sensors.
Preparing for disinfection or steri-run
1. Remove all samples from the workspace and store them in a safe place.
2. Procure a collecting vessel of sufficient capacity.
3. Place open end of hose into collecting vessel and lock drain hose valve safely into receptacle
4. The water reservoir starts draining.
Figure 8-1.
Drain valve of water reservoir
5. Allow the water reservoir to drain completely into the collecting vessel.
6. Wipe up residual water with a cloth.
7. Pull airbox (1/
) out of its seat on the water reservoir cover panel and remove it.
8. Take HEPA-filter (2/
) and reinstall the airbox.
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...