8
Cleaning and disinfection
Steri-run decontamination routine
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
8-9
Figure 8-4.
Decontamination routine phases
Activating steri-run
The steri-run decontamination routine is used to decontaminate the complete workspace.
1. Press the
steri-run
key.
• The steri-run menu is displayed.
Figure 8-5.
steri-run menu - Progress
2. steri-run menu - Exit and stop steri-run:
• Press the
End
key [3].
• The system returns to the main menu.
3. Activating steri-run:
• Press the START key.
• The steri-run-Progress dialog window is displayed.
4. To air the workspace, open both device doors when the audible alarm sounds after 30 seconds.
5. Remove all samples from the workspace.
6. Drain water from the reservoir, wipe up any residual water.
7. When the audible timer alarm is sounded, close both device doors.
Warming up
Decontamination
Cooling down
Start
End
Time
Temp.