Appendix C Scan/Calibration Procedure
C-8
25.
The Application gas liquid density default is "Off". Select "1=Change" to set a
value , or "
=Next".
26.
For Multi Gas Library Application Only: The Application gas GTWA weight
default is "Off". Select "1=Change" to set a value , or "
=Next".
27.
For Multi Gas Library Application Only: The instrument returns to the Gas
Summary screen for the second gas. Repeat steps 15-26 for each gas.
28.
For Single Gas Library Application Only: The Application wave count
default is 1. Select "
=Next".
For Multi Gas Library Application Only: Enter the number of wavelengths to
be analyzed.
29.
The Wavelength Summary screen shows the wavelength that the analyzer will
measure at. Select "1=Change".
30.
Again select "1=Change".
31.
Enter the wavelength recorded in step D-2, and select "Enter=Accept" (this
value MUST be entered!).
32.
The Wavelength H
2
O absorbance should be left at the default value to disable
humidity compensation. Select "
=Next".
33.
For Multi Gas Library Application Only: The instrument returns to the
Wavelength Summary screen for the second wavelength. Repeat steps 29-32
for each wavelength.
34.
The Application temperature compensation reference default value is "Off".
Select "
=Next".
Note:
To support temperature compensation, set this value to reference temperature
instead.
35.
The Application pressure compensation reference default value is "Off".
Select "
=Next".
Note:
To support pressure compensation, set this value to reference pressure instead.
36.
The Reference Compensation default is "Off". If the application requires a
reference wavelength, (4.004 is recommended), select "1=Change",
"1=Change", "2=Enabled" and "1=Change". Enter the reference wavelength
and "1=Change" to edit the application measurement drift (typically 2.0%).
"
=Next" (see “Reference Compensation” in Chapter 7 for an explanation of
this function). Select "Enter=Accept".
37.
The Humidity compensation default is "Off". Select "
=Next". Contact
Thermo Fisher Scientific before enabling humidity compensation for humidity
compensation coefficients.
Содержание MIRAN 205B Series
Страница 4: ......
Страница 14: ......
Страница 40: ......
Страница 54: ......
Страница 96: ......
Страница 102: ......
Страница 108: ......
Страница 112: ......
Страница 118: ......
Страница 122: ...Chapter 13 Infrared Spectroscopy Tutorial 13 2 Figure 13 1 Electromagnetic Spectrum ...
Страница 142: ......
Страница 170: ......
Страница 196: ......
Страница 202: ......
Страница 210: ......