Maintenance and Service
5-10
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the LCD assembly (
Figure 5–3
).
1.
Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2.
Unfasten the two captive hardware and swing cover down on the
power/com assembly.
3.
Disconnect the battery cable from J6 or J8 from the power
communication board.
4.
Unplug the ribbon cable from the LCD board.
5.
Unfasten the four nuts at the corners of the LCD board.
6.
Lift the LED screen out of the housing.
7.
Remove the four nuts from the LCD.
8.
Re-install the four nuts into the new LCD, and replace the LCD by
following the previous steps in reverse order.
CAUTION
Disconnect battery power and external power supplies before
servicing.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
LCD Assembly
Replacement
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......