Guidelines and Instrument Layout
Thermo Fisher Scientific
MIE ADR-1500 Instruction Manual
2-9
For ambient air sampling, the omni-directional inlet unit must be used to
minimize wind speed and direction effects on particle sampling
representativeness. This special omni-directional inlet is provided as a
standard accessory of the ADR-1500. On receipt by the customer, this inlet
will arrive packaged separately, and needs to be installed. Refer to
Figure 2–
7
for the final (installed) appearance of the omni-directional inlet. To
install proceed as follows:
Slide the omni-directional inlet onto the inlet stem of the ADR-1500 until
it bottoms. To remove the omni-directional inlet, lift and twist the unit
from the inlet stem.
If the ADR-1500 is to be used for extractive sampling (e.g., from a
chamber, duct, stack, etc.) a flexible plastic tubing (preferably electrically
conductive) can be used and connected to the inlet stem on the upper face
of the ADR-1500. In this case the omni-directional inlet is not used and a
3/8-inch compression fitting with Teflon or nylon ferrules should be used.
For sampling situations involving water sprays, fog, etc. it is recommended
that the in-line inlet heater be switched on. This optional operational
configuration ensures the sample relative humidity will not exceed 70%. If
not necessary, and to extend the run-time of the ADR-1500, when running
off of the internal battery, simply switch the internal heater off.
Equipment Damage
Damage may occur to the instrument if the
environmental conditions exceed IP65.
▲
Sampling
Guidelines
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......