Troubleshooting
6-8
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
The connector pin description in
Table 6–4
can be used along with the
board-level connection diagram to troubleshoot board-level faults.
Table 6–4.
ADR-1500 Communications Board Pin Out
Connector Label
Reference
Designator
Pin
Signal Description
24 VDC
J1
1
DC+
2
Power
Ground
Eternal 12/24 VDC
J2
1
DC+
2
Power
Ground
Pump In 0-8 VDC
Depending on flow
rate
J3 1
DC+
2
Power Ground (From Main Signal
Board)
Pump Out 0-8 VDC
Depending on flow
rate
J4 1
DC+
2
Power Ground (From Main Signal
Board)
Auxiliary/Radio 12VDC
J5
1
DC+
2
Power
Ground
Lead Acid Battery or
Charger VDC
J6 1
No
Connection
2
DC+
3
No
Connection
4
Power
Ground
USB J7
1
USB_VBUS
2
USB_D−
3
USB_D+
4
Ground
5
Ground
Chassis
Lead Acid Battery or
Charger VDC
J8 1
No
Connection
2
DC+
3
No
Connection
4
Power
Ground
Connector Pin
Descriptions
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......