Guidelines and Instrument Layout
2-12
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
There are three critical components to the bottom panel of the ADR-1500;
the universal AC power port (100-240 VAC 50/60 Hz), the auxiliary 12-
24 VDC power port, and the instrument exhaust. The AC power should be
connected whenever the internal batteries are exhausted or not present,
and/or when running continuously from the AC line. Any other DC source
(e.g., solar power supply, external battery, etc.) to be used to power the
ADR-1500 would be connected to its respective port. Please note that the
auxiliary 12-24 VDC will
not
charge the internal battery.
Refer to
Figure 2–6
below for the location of power port items on the
bottom of the ADR-1500.
Equipment Damage
Equipment damage may occur to instrument if power
inputs or fuse type exceeds specified ranges.
▲
Equipment Damage
Equipment damage may occur if exhaust port is
blocked or port covers are not in place if unused.
▲
Figure 2–6.
ADR-1500 Bottom Power Port
Bottom Power Port
Fuse Drawer
Exhaust
100-240 AC Input
50/60 Hz
12-24 VDC Input
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......