Guidelines and Instrument Layout
2-14
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
There are three features on the top of the ADR-1500; the inlet stem, the
beacon, and the carry handle. The inlet stem is used to draw the sample
into the nephelometric stage for aerosol sensing. This stem is 3/8-inch
O.D. and is compatible with 3/8 compression fittings. It is recommended
that the compression ferrules be made of nylon or Teflon; otherwise,
compressed steel ferrules may render the instrument useless for backward
compatibility with the intended inlet accessories. The connections to this
stem are:
●
Omni-directional inlet
●
Cyclone adapter
●
Zeroing filter
●
3/8-inch compression fitting with nylon or Teflon ferrules
The yellow beacon will flash if the alarm is enabled and the measured
concentration exceeds the threshold chosen by the user. The alarm may be
enabled for either an instantaneous value or for a STEL concentration.
The carry handle is to be used for carrying the instrument or hauling the
instrument to an elevated installation. The carry handle is designed for the
weight of the ADR-1500 only.
Equipment Damage
Do not attach additional items to the handle for
vertical hauling as this might compromise the strength of the handle and
the ADR-1500 enclosure.
▲
Refer to
Figure 2–8
below for the location of items on the inlet stem, the
beacon, and carry handle of the ADR-1500.
Figure 2–8.
ADR-1500 Top View
Top View
Inlet
Beacon
Carry Handle
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......