![Thermo Scientific MIE ADR-1500 Скачать руководство пользователя страница 115](http://html.mh-extra.com/html/thermo-scientific/mie-adr-1500/mie-adr-1500_instruction-manual_1105191115.webp)
Maintenance and Service
Thermo Fisher Scientific
MIE ADR-1500 Instruction Manual
5-27
Use the following procedure to replace the extended monitoring HEPA
filter (
Figure 5–11
).
1.
Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2.
Unfasten the two captive hardware and swing cover down on the center
panel.
3.
Disconnect the battery cable from J6 or J8 from the power
communication board.
4.
Use needle nose pliers and release the filter stem from the fittings.
5.
To install the new filter, follow the previous steps in reverse order.
Note
When replacing the new filter make sure filter arrow is facing up.
▲
Figure 5–11.
Replacing the Filters
Extended
Monitoring HEPA
Filter Replacement
In-Line Flow
Meter Filter
Extended Monitoring