Maintenance and Service
5-28
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the optional 37-mm filter cassette
holder assembly (
Figure 5–12
).
1.
Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2.
Unfasten the two captive hardware and swing cover down on the center
panel.
3.
Disconnect the battery cable from J6 or J8 from the power
communication board.
4.
Remove the filter by depressing the overlapping push connect fitting to
release the HEPA filter.
5.
To install the new filter, follow the previous steps in reverse order.
Figure 5–12.
Replacing the 37-mm Filter Cassette Holder Assembly
37-mm Filter Cassette
Holder Assembly
Replacement
(Optional)
37-mm Filter
Cassette Holder