Troubleshooting
Thermo Fisher Scientific
MIE ADR-1500 Instruction Manual
6-5
During zeroing, the ADR-1500 also performs diagnostics on the optics and
flow rates. The only allowable fault is “Background High”. All other faults
will result in a failure to zero the instrument. All faults are reported with a
hex coded error flags. The faults can be determined from the following
Table 6–2
, but if a fault persists, the ADR-1500 instrument requires
service.
Table 6–2.
Troubleshooting – Error Code Responses from Zeroing
Hex code [Flags]
Binary
Fault
01 x
0000 0001 b
Background High
02 x
0000 0010 b
Background calculation FAILED
04 x
0000 0100 b
Source current low
08 x
0000 1000 b
Source current high
10 x
0001 0000 b
Ref. detector low
20 x
0010 0000 b
Ref. detector high
40 x
0100 0000 b
Pump voltage low
80 x
1000 0000 b
Pump voltage high
Instrument
Status Flags