Maintenance and Service
5-20
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the heater switch assembly (
Figure
5–8
).
1.
Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2.
Unfasten the two captive hardware and swing cover down on the
power/com assembly.
3.
Disconnect the battery cable from J6 or J8 from the power
communication board.
4.
Unplug the connector from board “HEATER SW”.
5.
Use needle nose pliers and squeeze the sides and push switch out of
cover.
6.
Snap the new replacement switch into cover housing (see
Figure 5–9
).
Make sure the 1 is facing left and the 0 is facing right. Refer to the
“ISO” view in
Figure 5–9
.
7.
Connect wire harness “lug end” to the tope 2 or bottom 2, but
not one
of each
. Connect red wire to gold terminal and black wire to silver
terminal.
8.
Connect the other end of the wire harness to “HEATER SW” J17 on
the board.
9.
Replug the power by following steps 1 to 3 in reverse order.
Heater Switch
Assembly
Replacement
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......