Maintenance and Service
Thermo Fisher Scientific
MIE ADR-1500 Instruction Manual
5-19
15.
Unfasten the four captive hardware, and slide whole enclosure assembly
downwards and out.
16.
Send the entire optic enclosure assembly to Thermo Fisher Scientific
for repair or maintenance.
17.
To facilitate reconnection, repeat steps 4 to 13 in reverse order and
refer to
Figure 5–6
.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲