Maintenance and Service
5-24
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the lead acid battery (
Figure 5–10
).
1.
Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2.
Unfasten the two captive hardware and six screws to fully remove the
cover.
3.
Disconnect the battery cable from J6 or J8 from the power
communication board.
4.
Remove the battery cover plate and slide out the old battery.
5.
Insert new battery, and follow previous steps in reverse order.
CAUTION
Disconnect battery power and external power supplies before
servicing.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
Equipment Damage
Replace with specified battery only.
▲
Lead Acid
Battery
Replacement