Thermo Fisher Scientific
MIE ADR-1500 Instruction Manual
B-1
Appendix B
Serial Commands
This appendix provides a list of the ADR-1500 Serial Port Commands that
can be used to remotely control the instrument.
Connect to the ADR-1500 at 19200, n, 8, 1, no flow control. The USB
connection is actually a virtual serial port and will request appropriate
drivers the first time it fires up in a windows environment. These are
available with the pDR Port install disk.
The protocol is designed for human terminal access. Commands are
executed when the user hits Enter. Spaces between tokens do not matter.
Generally, parameters not included in the command line are defaulted.
First is a list of key worded commands that should be made available to the
end user as a public interface. With these commands you can completely
control the instrument.
ALARM
[
status
] [
alarm concentration level
]
status=
{0, 1, 2,}
0 = DISABLED
1 = INSTANT
2 = STEL
alarm concentration level
is specified in mg/m
3
(0.01 to 400 [mg/m
3
])
ANALOGOUT
[
analog output
menu choice
]
If the units are mass concentration (mg/m
3
), the
analog output menu choices
= {0, 1, 2, 3, 4, 5, 6, 7, 8, 9}
0 = DISABLED
1 = 0.10 mg/m
3
2 = 0.40 mg/m
3
3 = 1.00 mg/m
3
4 = 4.00 mg/m
3
5 = 10.0 mg/m
3
6 = 40 mg/m
3
7 = 100 mg/m
3
8 = 400 mg/m
3
9 = 1000 mg/m
3
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......