Optional Accessories
8-4
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Note
Only necessary terminals are present on single throw models.
▲
Each of the above two switches can switch loads of up to 30 amperes.
Once the leads have been connected to the appropriate terminals of the
alarm relay, the sheet metal cover should be reinstalled and secured. The
leads should be passed through the adjacent free feed-through cable ground
and connected as a switch to apply power to the required external
equipment to be controlled by the alarm condition. For DC power
application, connect the wire from terminal #1 to the negative terminal of
the power source (i.e., battery or D.C. supply). Finally, plug in the cyclone
making sure to insert it completely.