Operation
3-14
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
The Configure Menu is used to access the list of displays allowing the user
to configure how the instrument measures and logs data including “Display
Avg Time”, “Flow Rate”, “Logging Parameters”, “Inlet Type”, “RH
Correction”, “Alarm”, “Units”, “Analog Output”, “LCD Backlight”, LCD
Contrast” and “Time/Date”.
CONFIGURE
The Display Average Time screen offers the user to adjust the running
(boxcar) average time of the displayed concentration (or scattering
coefficient) between 1 and 60 seconds. The average concentration displayed
is the value scaled to the analog output and offered as an alarm criterion.
●
From the Configure menu, press the
ENTER
key and scroll the
▲
or
▼
key to the Display Avg Time screen.
DISPLAY AVG TIME
10 seconds
The Display Average Time Edit screen allows the user to change the
running average displayed on the screen between 1 and 60 seconds. This
also doubles as the running average that is provided as an analog output
and also as one of the values that triggers the alarm. Use the
▲
and
▼
keys
to increase or decrease the display average time by seconds. The asterisk (*)
appears to the left of the field being edited.
●
From the Display Avg Time screen, press the
ENTER
key and edit the
display averaging time.
●
Press the
ENTER
key to save the display averaging time and return to
Display Avg Time screen.
DISPLAY AVG TIME
* 11 seconds
Configure Menu
Display Average
Time
Display Average Time
Edit
Содержание MIE ADR-1500
Страница 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Страница 4: ......
Страница 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Страница 26: ......
Страница 105: ...Maintenance and Service Thermo Fisher Scientific MIE ADR 1500 Instruction Manual 5 17 Figure 5 7 Wiring Diagram 2 ...
Страница 132: ......
Страница 140: ......