Outputs and Alarm
7-4
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
The voltage alarm signal is available between pins # 1 through 5 on the
ANALOG OUTPUTS connector in the upper middle of the side panel. In
Figure 7–1
below, the alarm driver circuit is depicted.
Figure 7–1.
ADR-1500 External Analog Driver
Table 7–1.
Analog Output Descriptions
Analog Out
Description
Wire
Pin 1
Voltage_OUT
RED
Pin 2
Current_OUT
BLUE
Pin 3
Power Ground
BLACK
Pin 4
Alarm
ORANGE
Pin 5
Power Ground
GREEN
WARNING
Do not apply AC voltage to this connector. Maximum
allowable DC voltage is 30 V.
▲
Analog Outputs