Troubleshooting
6-4
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Malfunction
Possible Cause
Action
No USB
communication
Wrong COM port (pDR Port)
Verify COM port (pDR Port).
Disconnected cable
Verify cable.
Defective external USB cable
Replace cable.
Defective internal USB cable
Send in for service.
Non-working strobe
Disconnected cable
Verify cable connection.
Defective strobe
Send in for service.
No relay (if installed)
Disconnected cable
Verify cable connection.
Defective relay
Replace relay.