Dionex AutoTrace 280 Operator’s Manual
64
Doc. 065330-05 03/2020
•
Load the sample rack.
3.9.2
Running Your Method
•
Load the method.
•
Intervene as required.
•
Press
CONT
to silence the end-of-run alarm.
3.9.3
After Running the Method
•
Remove the used cartridges or disks.
•
Prepare to run the next batch of samples by cleaning the sample lines
(see
•
If done for the day, clean up by emptying the waste containers and
flushing the solvent lines (see
3.10 Routine Maintenance
This section describes routine maintenance procedures that users can perform. All
other Dionex AutoTrace 280 maintenance procedures must be performed by
Thermo Fisher Scientific personnel.
In order to test, adjust, or replace components not covered in this section, run the
Dionex AutoTrace 280 diagnostic methods. For details, see
.
3.10.1 Daily Maintenance
•
Check the waste containers and empty as needed.
•
Flush the solvent lines (see
•
Flush the sample lines (see
Acetone is harmful to the sample pumps and should never be used to
clean the sample path or rinse the sample container.
Do not run pumps dry. Always leave deionized water in the sample
lines when the system is not in use.
Содержание Dionex AutoTrace 280
Страница 1: ...Dionex AutoTrace 280 SPE Instrument Operator s Manual Document No 065330 Revision 05 March 2020...
Страница 86: ...Dionex AutoTrace 280 Operator s Manual 78 Doc 065330 05 03 2020...
Страница 134: ...Dionex AutoTrace 280 Operator s Manual 126 Doc 065330 05 03 2020...
Страница 158: ...Dionex AutoTrace 280 Operator s Manual 150 Doc 065330 05 03 2020...