Dionex AutoTrace 280 Operator’s Manual
88
Doc. 065330-05 03/2020
Remove the Defective Syringe
1. Remove the two screws from the mounting bracket that secure the syringe to
the housing (see
Figure 5-2. Syringe Screws and Mounting Bracket
2. Remove the black thumbscrew from the bottom of the syringe plunger.
3. Gently slide the syringe assembly out of the housing.
4. Loosen the small black thumbscrew on the mounting bracket.
Housing
Fitting
Screw
Thumbscrew
Plunger
Driving Arm
Black Thumbscrew
Black Mounting
Bracket