Dionex AutoTrace 280 Operator’s Manual
56
Doc. 065330-05 03/2020
When you complete the manual steps, press
CONT
to have the system continue
running the method.
3.6.1
Pausing a Run
To suspend operation at any time, press
PAUSE
. The system will
complete the function in process when
PAUSE
was pressed, and then
stop.
If you press the
PAUSE
button before all method steps have finished
running, be aware that pressing the
CONT
button later (to restart the
method) will resume operation
at the point when processing was
interrupted
, not at step #1 of the method.
3.6.2
Resuming a Run
To resume running a method after pressing
PAUSE
, press
CONT
.
If you press the
PAUSE
button before all method steps have finished
running, be aware that pressing the
CONT
button to restart the method
will resume operation
at the point when processing was interrupted
, not at
step #1 of the method.
Concentrate Sample
CONCENTRATION STEP
PLACE EMPTY CARTRIDGES
IN THE HOLDERS
PRESS CONT TO RESUME
Clean Sample Path
TO CLEAN SAMPLE PATH
PLACE SAMPLE INPUT
LINES INTO A BEAKER
PRESS CONT TO RESUME
Pause and Alert
SYSTEM PAUSED
PRESS CONT TO RESUME
Method Step
Front Panel Display
Table 3-1. Front Panel Messages (Continued)
Содержание Dionex AutoTrace 280
Страница 1: ...Dionex AutoTrace 280 SPE Instrument Operator s Manual Document No 065330 Revision 05 March 2020...
Страница 86: ...Dionex AutoTrace 280 Operator s Manual 78 Doc 065330 05 03 2020...
Страница 134: ...Dionex AutoTrace 280 Operator s Manual 126 Doc 065330 05 03 2020...
Страница 158: ...Dionex AutoTrace 280 Operator s Manual 150 Doc 065330 05 03 2020...