Dionex AutoTrace 280 Operator’s Manual
92
Doc. 065330-05 03/2020
2. Press
CONT
to run the method.
Remove the Defective Syringe
1. Loosen the knurled knobs on both sides of the syringe clamp. Pivot the open
end of the syringe clamp out of the way (see
Figure 5-5. Syringe Clamp Knurled Knobs
2. Remove the black thumbscrew from the bottom of the syringe plunger.
3. Gently slide the syringe assembly out of the housing.
Luer-Lock Fitting
Syringe Clamp
Housing Tab
Knurled Knob
Chrome
End Piece
Syringe Plunger
Driving Arm
Thumbscrew