5 • Service
Doc. 065330-05 03/2020
91
Figure 5-4. Thumbwheel Rotation
5. Tighten the knurled thumbscrew to secure the position.
Complete the Procedure
Run the Benchmark Test diagnostic method (#30) (see
) to verify that
the new syringe works properly.
5.6
Replacing the Air Push Syringe
Parts Needed
•
Air push syringe, 10 mL (P/N 070296)
Center the Syringe Plunger
Follow these steps to center the syringe plunger and prepare the syringe for
removal.
1. Select the Adjust Air Syringe diagnostic method:
a.
Press
LOAD
to display the
LOADING METHOD
screen.
b. Press
LOAD
to increment the displayed method to 27.
c.
Press
CONT
once to select method 27.
Thumbwheel
Top View
Turn thumbwheel one
more partial turn to the
right by the amount shown.