90
Doc. 065330-05 03/2020
Figure 5-3. Syringe Knurled Thumbscrew
2. Turn the thumbwheel to the left two full turns.
3. Select the Adjust Liq Syringe diagnostic method (#26) again and press
CONT
to run the method. As the plunger moves up, you should hear a clicking
sound.
4. Turn the thumbwheel to the right until the clicking just stops. Then turn it one
more partial turn (see
Fitting
Thumbwheel
Driving Arm
Plunger
Knurled
Thumbscrew