Dionex AutoTrace 280 Operator’s Manual
118
Doc. 065330-05 03/2020
Parts Needed
•
Six 16 x 100 mm test tubes and elution rack
–or–
For alignment of the 40 mL elution rack only: Six 40 mL collection vials and
elution rack
•
Sample container with at least 100 mL of water
•
Solvent containers filled with colored water
Run the Benchmark Test
1. Select the Benchmark Test diagnostic method:
a.
Press
LOAD
to display the
LOADING METHOD
screen.
b. Press
LOAD
to increment the displayed method to 30.
c.
Press
CONT
once to select method 30.
2. Press
CONT
to run the method.
3. As the method runs, follow the on-screen instructions to verify that the:
•
Front panel
RUN
LED turns on.
•
Front panel
ERROR
LED turns on.
•
Elution shuttle moves to its three different positions.
•
Exhaust fan turns on.
•
Cartridge holder’s LEDs are lighted.
4. Perform the following wet test whenever tubing, valves, or syringes have
been replaced. Follow the on-screen instructions and do the following:
a.
Place six empty test tubes in the 16 x 100 mm elution rack.
–or–
Place six empty 40 mL collection vials in the 6 x 40 elution rack.
b. Place the system’s five solvent lines in containers holding water solutions
of different colors.
c.
Observe that the tubes are filled with water as follows:
Tube #1
fills with 10 mL from
solvent container #1
.
Содержание Dionex AutoTrace 280
Страница 1: ...Dionex AutoTrace 280 SPE Instrument Operator s Manual Document No 065330 Revision 05 March 2020...
Страница 86: ...Dionex AutoTrace 280 Operator s Manual 78 Doc 065330 05 03 2020...
Страница 134: ...Dionex AutoTrace 280 Operator s Manual 126 Doc 065330 05 03 2020...
Страница 158: ...Dionex AutoTrace 280 Operator s Manual 150 Doc 065330 05 03 2020...