3 • Operation and Maintenance
Doc. 065330-05 03/2020
49
If...
Then...
You are going between immiscible
solvents in the method
Perform a
Wash Syringe
step
with an intermediate solvent
before performing the next step
(even if
Autowash
is being
used)
You need syringe washing to appear in
your method printout for auditing
purposes
Disable
Autowash
and use
Wash Syringe
method steps
where needed
You are running with the
Air Factor
parameter set to less than one
Perform a
Cartridge Rinse
(or
Disk Rinse
) step between
sample loading and elution
The elution solvent is immiscible in water
Perform a
Dry with Gas
step to
remove all water from the
cartridges or disks and eliminate
channeling
You are not drying the cartridge or disk
after loading sample
Perform a
Cartridge Rinse
(or
Disk Rinse
) step before elution
There is water in your extract
Increase the pressure and/or time
of the
Dry with Gas
step
The elution solvent is not reaching the
packing material in the first
Elute to
Soak
step
Use a volume of 3 mL or greater
in the elution step
You need to increase the contact time
between solvent and packing material for
extraction of analytes
Use an
Elute to Soak
step
before
Elute to Collect
You need to minimize the amount of air
introduced between multiple elution steps
Use an
Elute to Soak
step
before
Elute to Collect
You need to collect more than the
maximum elution volume of your
collection container
Use an
Elute to Collect, 2nd
Tube
step to double the
collection volume
You need to improve throughput
Decrease times wherever
possible
You need to improve recovery
See
Table 3-3. Method Step Considerations (Continued)
Содержание Dionex AutoTrace 280
Страница 1: ...Dionex AutoTrace 280 SPE Instrument Operator s Manual Document No 065330 Revision 05 March 2020...
Страница 86: ...Dionex AutoTrace 280 Operator s Manual 78 Doc 065330 05 03 2020...
Страница 134: ...Dionex AutoTrace 280 Operator s Manual 126 Doc 065330 05 03 2020...
Страница 158: ...Dionex AutoTrace 280 Operator s Manual 150 Doc 065330 05 03 2020...