5 • Service
Doc. 065330-05 03/2020
89
5. While holding the black mounting bracket in your left hand and the syringe in
your right hand, turn the syringe counterclockwise until it disengages from the
fitting.
6. Separate the syringe from the fitting.
Mount the New Syringe
1. Slide the new syringe into the fitting. When aligned, turn the syringe
clockwise to secure it in the fitting.
2. Tighten the small black thumbscrew on the mounting bracket.
3. Slide the syringe assembly into the housing and secure the mounting bracket,
using the two screws previously removed.
4. Gently pull the syringe plunger down until it contacts the top of the driving
arm.
5. Secure the plunger to the arm with the black thumbscrew previously removed.
Adjust the New Syringe
1. Loosen the knurled thumbscrew approximately two turns (see