5 • Service
Doc. 065330-05 03/2020
83
8. Install the fittings into the ports of the component. Reinstall the component, if
it was removed for easier access. Tighten fittings securely to prevent leaks.
Tip:
If necessary, use latex gloves to obtain a tight grip.
9. If you removed the Dionex AutoTrace 280 rear panel, reinstall the rear panel
with the six screws previously removed.
10. Reconnect the power cord and turn on the power.
11. Place solvent lines in the solvent reservoirs.
12. Run the Benchmark Test diagnostic method (#30) (see
5.2
Tightening the Nitrogen Manifold Fittings
If there is a leak from the nitrogen manifold, tighten the fittings.
Parts and Tools Needed
•
#2 Phillips screwdriver
•
Soap solution
•
Cotton-tipped swabs
Tighten the Fittings
1. Turn off the Dionex AutoTrace 280 power switch and disconnect the power
cord from both its source and from the Dionex AutoTrace 280 right-side
panel.
2. Using the #2 Phillips screwdriver, remove the six Phillips screws securing the
rear panel to the system. Set the panel and screws aside.
HIGH VOLTAGE—Disconnect the main power cord from its source, as
well as from the right-side panel of the Dionex AutoTrace 280.
HAUTE TENSION—Débranchez le cordon d'alimentation principal de
sa source et du panneau de droit-côté du Dionex AutoTrace 280.
HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose und
der Netzbuchse auf dem rechten Seitenteil des Dionex AutoTrace 280.
Содержание Dionex AutoTrace 280
Страница 1: ...Dionex AutoTrace 280 SPE Instrument Operator s Manual Document No 065330 Revision 05 March 2020...
Страница 86: ...Dionex AutoTrace 280 Operator s Manual 78 Doc 065330 05 03 2020...
Страница 134: ...Dionex AutoTrace 280 Operator s Manual 126 Doc 065330 05 03 2020...
Страница 158: ...Dionex AutoTrace 280 Operator s Manual 150 Doc 065330 05 03 2020...