3 • Operation and Maintenance
Doc. 065330-05 03/2020
55
3.6
Running a Method
To run a Dionex AutoTrace 280 method:
1. Press
LOAD
until the required method number is displayed.
2. Press
CONT
to select the displayed method.
3. Press
CONT
again to start running the selected method.
While the method runs, the Dionex AutoTrace 280 front panel reports each step as
it is performed, as well as the time remaining until the step is completed, if
applicable. For example, while a sample is being loaded (pumped) onto the
cartridge, the front panel might display:
LOADING SAMPLES
ONTO CARTRIDGE
@ 10 MLS/MIN
5.0 Minutes Left
NOTE When the installed system is the Dionex AutoTrace 280
disk model, “Disk” replaces “Cartridge” in all messages
displayed on the front panel.
The display also reports when manual intervention is required and when the
sample run is finished.
Some method steps cause the system to pause operation and wait for manual
intervention. When this happens, the system beeps (if the
Beeper On
check box in
the Setup Parameters dialog box is selected) and the front panel displays a
message indicating the required action.
lists these front panel messages.
Method Step
Front Panel Display
Rinse Sample Container
RINSE SAMPLE
CONTAINER WITH
xx ML OF SOLVENT
PRESS CONT TO RESUME
Elute to Collect, 2nd Tube
SECOND ELUTION STEP
REPLACE TUBES
WITH NEW TUBES
PRESS CONT TO RESUME
Table 3-1. Front Panel Messages
Содержание Dionex AutoTrace 280
Страница 1: ...Dionex AutoTrace 280 SPE Instrument Operator s Manual Document No 065330 Revision 05 March 2020...
Страница 86: ...Dionex AutoTrace 280 Operator s Manual 78 Doc 065330 05 03 2020...
Страница 134: ...Dionex AutoTrace 280 Operator s Manual 126 Doc 065330 05 03 2020...
Страница 158: ...Dionex AutoTrace 280 Operator s Manual 150 Doc 065330 05 03 2020...