Figure 4-1. Multi-Point Calibration Page
4.2.1.1
ZERO CALIBRATION CHECK AND ACTUAL CALIBRATION
1. Go to the Calibration>M-P menu.
2. Input Zero air through the Sample port and press the Start button.
3. Either check or calibrate as follows:
CHECK ONLY:
ACTUAL CALIBRATION:
a. Wait for reading to stabilize.
b. Press Stop and check the
reading.
a. Press the Zero button.
b. Press Stop and check the reading.
4.2.1.2 SPAN CALIBRATION CHECK AND ACTUAL CALIBRATION
1. While still in the Calibration>M-P menu, input Span gas through the Sample
port and press the Start button.
2. Either check or calibrate as follows:
CHECK ONLY:
ACTUAL CALIBRATION:
a. Wait to reach stability,
then press Stop.
b. Record the reading(s).
a. Press the Set Span Target button and
enter the span gas concentration.
b. Verify the concentration reading is the
same as the span gas concentration
being supplied.
c. If correct, wait to reach stability, then
press the Span button.
d. In the Cal Result window, press OK.
3. Press the Stop button and return to Home screen.
4. In the Dashboard, check and record the Slope(s) and the Offset(s). (See
Section 4.4 regarding these values).