2.3.2.4 PNEUMATIC CONNECTIONS: AMBIENT AIR MONITORING IN SAME ROOM
In this application it is often preferred that the sample gas and the source gas for the O
3
generator and internal zero air be the same chemical composition.
Enclosure Wall
Instrument Chassis
(with IZS installed)
ZERO AIR
DRY AIR
SAMPLE
EXHAUST
SPAN
Minimum 3 ft. in length
SAMPLE
and
DRY AIR
inlets
open to room
1
1
1
Figure 2-16. Gas Line Connections when the T400 Analyzer is Located in the Room Being Monitored
S
AMPLE
G
AS AND
D
RY
A
IR
S
OURCES
For instruments located in the same room being monitored, there is no need to attach the
gas inlet lines to the
SAMPLE
inlet or the
DRY AIR
inlet.
E
XHAUST
O
UTLET
Attach an exhaust line to the EXHAUST outlet fitting. The end of the exhaust line
should be at least 2 feet from the rear panel of the instrument in order to prevent sampling
its own exhaust (resulting in low readings).
L
EAK
C
HECK
Once the appropriate pneumatic connections have been made, check all pneumatic
fittings for leaks using the procedures defined in Section 5.6.3.