2.3.1.5 CONNECTING THE CONTROL INPUTS (DIGITAL INPUTS)
With zero and span valves option installed, their digital control inputs are provided
through the rear panel 10-pin
CONTROL IN
connector
to remotely activate the zero and
span calibration modes.
Energize the Control Inputs either by the in5V available from the pin labeled “+”
(more convenient), or by a separate external 5 VDC power supply (ensures that these
inputs are truly isolated). Refer to Figure 2-7
CONTROL IN
A B C D E F U
+
LO
S
PA
N
Z
E
R
O
S
PA
N
CONTROL IN
A B C D E F U
+
LO
S
PA
N
Z
E
R
O
S
PA
N
-
+
5 VDC Power
Supply
Local Power Connections
(non-isolated configuration)
External Power Connections
(isolated configuration)
Figure 2-7. Control Input Connector
Table 2-5. Control Input Signals
INPUT #
STATUS DEFINITION
ON CONDITION
A
REMOTE ZERO CAL
The analyzer is placed in zero calibration mode.
B
REMOTE LOW SPAN CAL
The analyzer is placed in low span calibration mode.
C
REMOTE SPAN CAL
The analyzer is placed in span calibration mode.
D, E & F
SPARE
Digital Ground
The ground level from the analyzer’s internal DC power supplies (same as
chassis ground).
U
External Power input
Input pin for +5 VDC required to activate pins A – F.
+
5 VDC output
Internally generated 5V DC power. To activate inputs A – F, place a jumper
between this pin and the “U” pin. The maximum amperage through this port
is 300 mA (combined with the analog output supply, if used).