6.3.3.5
THERMOCOUPLE INPUTS AND CONFIGURATION JUMPER (JP5)
In its base configuration, the T400 analyzer does not include any thermocouple sensors;
however, in instruments where the optional metal wool scrubber (OPT-68) is installed,
one thermocouple is used to sense the temperature of the scrubber. By default, this single
thermocouple input is plugged into the TC1 input (J15) on the relay PCA. TC2 (J16) is
currently not used.
Table 6-2. Thermocouple Configuration Jumper (JP5) Pin-Outs
TC INPUT
JUMPER
PAIR
DESCRIPTION
FUNCTION
TC1
1 – 11
Gain Selector
Selects preamp gain factor for J or K TC
OUT = K TC gain factor
;
IN = J TC gain factor
2 – 12
Output Scale Selector
Selects preamp gain factor for J or K TC
OUT = 10 mV / °C;
IN = 5 mV / °C
3 – 13
Type J Compensation
When present, sets Cold Junction Compensation for J
type Thermocouple
4 – 14
Type K Compensation
When present, sets Cold Junction Compensation for K
type Thermocouple
5 – 15
Termination Selector
Selects between Isolated and grounded TC
IN = Isolate TC;
OUT = Grounded TC
TC2
NOT USED