58
Teledyne API – Model T400 O
3
Analyzer
083730400A DCN8130
2.4.5 U
TILITIES
The Utilities menu has a variety of functions as described next in Table 2-9.
Table 2-9. Utilities Submenu Descriptions
UTILITIES
MENU
DESCRIPTION
Datalog View
Displays the data logs that were configured via the Setup>Data Logging menu. From this list
a log can be selected and filters applied to view the desired data. (For details on setting up
and running the Data Logger, see Section 2.5.1).
Alerts Log
Displays a history of alerts that are triggered by factory-defined and user-defined Events,
such as warnings and alarms (See Section 2.5.2 for Events configuration).
USB Utilities
Serves multiple purposes using a flash drive connected to the instrument’s front panel USB
port:
•
download data from the instrument’s Data Acquisition System (DAS), the Data Logger, to
a flash drive (Section 2.5.1.3)
•
•
download a basic operation functionality report (Section 5.3).
Diagnostics
Provides access to various pages that facilitate troubleshooting.
Analog Outputs
Show the voltage signals for the functions selected and configured in the
Setup>Analog Outputs menu (Section 2.5.8).
Digital Inputs
Show whether specific available features are turned ON or OFF; for
example, whether or not Maintenance Mode input or Language selection
can be controlled through the front panel, or whether a zero or span
calibration can be activated remotely when an external source is
connected to the rear panel Control In connector.
Digital Outputs
Show the function of user-specified parameters configured in the
Setup>Digital Outputs menu (Section 2.5.7).
Flow Cal
Use to calibrate the sample gas flow reading with actual flow measured
by an external device. (See Section 4.5.2).
O3 Gen Cal
Use to calibrate the O3 Generator option
Pressure Cal
Use for compensating for changes in atmospheric pressure
(Section 4.5.1).
Config Transfer
Use to transfer instrument configuration from/to other same-model
instruments (Section 2.6)
2.4.6 S
ETUP
The Setup menu is for configuring the instrument’s various programmable features and
functions. Section 2.5 provides details for the menus under Setup.