083730400A DCN8130
Teledyne API – Model T400 O
3
Analyzer
133
Figure 5-14. Critical Flow Restrictor Assembly/Disassembly
5.7.15.2 REPLACING THE STANDARD REFERENCE O3 SCRUBBER
To determine whether the reference O
3
scrubber requires replacement, follow the
procedures in Section
Error! Reference source not found.
.
1. Turn off power to the instrument.
2. Remove instrument cover.
3. The reference scrubber is a blue colored canister located at the rear of the
measure/reference valve assembly. See
Error! Reference source not found.
.
4. Disconnect the top 1/8” brass tube fitting from the scrubber.
5. Carefully remove the scrubber from the retaining clip.
6. Remove the bottom 1/8” brass tube fitting from the scrubber.
7. Perform the above steps in reverse to install the new scrubber.
8. Run the instrument for at least 24 hours with the new scrubber installed, and
recalibrate the instrument.
5.7.15.3 REPLACING THE IZS O
3
SCRUBBER
1. Turn off power to the instrument.
2. Remove instrument cover.
3. The IZS zero air scrubber is attached to the brass elbow inlet fitting on the top of the
O
3
generator assembly. See Figure 5-15.
4. Disconnect 1/4” Tube Fitting nut on O
3
generator inlet fitting.